Seq. No. 1

CALIBRATION CERTIFICATE

Customer

Rahul Patel
Plot No. 7+8,Khare Group Hsg,Project,Old Dhamani, Golden Circle, Surat, 355 876

Certificate No.
UE/SLI/000178.0/24

Calibration Date : 01-05-2023

Certificate Date : 01-01-1970

Recommended Calibration Due Date : 01-11-2023

Nomenclature : BLOOD BANK REFRIGERATOR

Location : BLOOD CENTRE

User ID : ABC1

Make : XYZ1

Sr.No. : 1111

Equipment Range : From: 0 °C To: RT

Operating Range : From: 2 °C To: 6 °C

Discipline : THERMAL

Parameter : TEMPERATURE

Date of Receipt : 01-05-2023

Status of Equipment on receipt : WORKING

Extra Attribute :

Extra Attribute :


Environmental conditions

Ambient Temp. : 25±4°C

Relative Humidity : 70%RH

Calibrated ' At Site ' at Customers address

STANDARD USED

1) Name : DIGITAL HUMIDITY INDICATOR Make/Model : MAXTECH Cal At : ERTL, MUMBAI Sr No : UE/CS/A/STD/04 Cal Date : 27-05-2022 Valid upto : 27-05-2023 Cert No. 2022TNP47

2) Name :MULTIFUNCTION CALIBRATOR Make/Model : FLUKE Cal At : IDEMI, MUMBAI Sr No : UE/CS/A/STD 01 Cal Date : 18-05-2022 Valid upto : 17-05-2023 Cert No. CC/ECL/0227/22-23

3) Name :PT-100 SENSOR WITH TIC Make/Model : UDHYAMSHRI Cal At :IDEMI, MUMBAI Sr No :UE/CS/A/TSD/01 Cal Date :13-05-2022 Valid upto : 12-05-2023 Cert No. CC/TRL/0039/22-23

4) Name :PT-100 SENSOR WITH TIC Make/Model : UDHYAMSHRI Cal At :IDEMI, MUMBAI Sr No :UE/CS/A/TSD/01 Cal Date :13-05-2022 Valid upto : 12-05-2023 Cert No. CC/TRL/0039/22-23

5) Name :PT-100 SENSOR WITH TIC Make/Model : UDHYAMSHRI Cal At :IDEMI, MUMBAI Sr No :UE/CS/A/TSD/01 Cal Date :13-05-2022 Valid upto : 12-05-2023 Cert No. CC/TRL/0039/22-23

* All master calibrator calibration certificates are traceable to National standards with unbroken chain.


Calibration Procedure No.:UE-SOP-T-001

Observations

Sr. No. DUC Reading (°C) Mean Standard Reading (°C) Uncertainty (±) (°C)
1 4.2 4.42 0.26
2 3.15 3.50 0.26
3 5.23 5.44 0.26
4 4.25 4.41 0.25
5 5.62 5.46 0.25
6 4.2 4.42 0.26
7 3.15 3.50 0.26
8 5.23 5.44 0.26
9 4.25 4.41 0.25
10 5.62 5.46 0.25

DUC = Device Under Calibration

(Uncertainty of the measurement is stated in the result at 95% Confidence Level for k=1.96)